opened from the file open menu or button, or uploaded from an instrument connected to the
PC via the configuration socket or serial communications module. The new profile can be
download to the instrument or saved to disk with a .pfl file extension.
New / Open / Save / Print Upload Profile Download Profile Mode Drop-Down
If the option to uploaded a profile is chosen, a list of profiles in
the connected instrument is shown. The user can select the
required profile from the list.
A directory of existing profiles in the instrument can also be
requested. This allows one or all of the profiles to be deleted.
When downloading a profile to the instrument via the
configuration socket or over serial communications, a list of
existing profiles and empty profile slots is displayed. The user
can select where to place the profile (a warning is shown if the
profile will overwrite an existing profile).
The number of free segments still available is also shown.
A drop-down menu in the Profile Editor switches between the Profile Header and Segment
Data. Refer to the Profiler Setup Menu and Profiler Option sections for full details of the
header and segment data.
Header data includes a 16-character profile name, options for starting the profile after a delay
or at a specific day and time, the starting setpoint, the action to take after a power failure or
profile abort and the number of times the profile will run.
The segments are shown in Segment Data mode. The last segment is always an End, Join
or Repeat Sequence type, and cannot be deleted. The user can select and change any
segment’s type and values, and they can insert additional segments before the selected one.
A dynamically scaled graphic shows the segments of the profile, with the select segment is
highlighted in red. The five profile events are shown below the graph.
Insert Segment Un-Zoom Profile Directory Selected Segment (Red)
Changing the Start-up Splash Screen
The graphic shown during the instrument start-up sequence can be changed from the main
parameter screen. Select Download Splash Screen option from the Device menu. Choose
your new graphic file (most common file types are supported). The chosen image will
converted to monochrome and be rescaled to 160 pixels wide by 80 pixels high. For best
results, the image should be simple and have an aspect ration of 2:1. Complex graphics with
multiple colours or greyscales will not reproduce well. A preview of the results is shown.
Click the Download button to store it to the instrument.
Changing the Alternate Display Language
The alternate language can be changed from the main parameter screen. Select the
Download Language File option from the Device menu. Choose the correct file (language
files have a .bin extension) and click the Open button to store it to the instrument.
Instrument Simulation
A fully functional and interactive instrument simulation is included with the software. This is
linked to a simulated process, allowing changes to a configuration to be tested before use.
Figure 58. Honeywell DCP200 Simulation
Configuring The Connection
|