详细介绍:
C25TCOUA1100
联系人:魏工 电话:15882174684
座机:028-84472033
QQ:2030149354
产品资料:
Sensor Break
Protection:
Break detected within two seconds. Process Control outputs go to the
pre-set power value. High and Senor Break Alarms operate.
Resistance Temperature Detector (RTD) Input
RTD Types & Ranges
Sensor
Type
Range in °C Range in °F Sensor
Type
Range in °C Range in °F
3-Wire
PT100
-199 to 800°C -328 to 1472°F NI120 -80 to 240°C -112 to 464°F
Note:
The Scale Range Upper Limit and Scale Range Lower Limit parameters, can be used to
restrict range. An optional decimal place can be displayed up to 999.9°C/F
RTD Performance
Measurement
Accuracy:
±0.1% of span ±1LSD.
Linearization
Accuracy:
Better than ±0.2°C any point (±0.05°C typical).
PT100 Input complies with BS1904 and DIN43760 (0.00385Ω/Ω/°C).
Sensor Resistance
Influence:
Pt100 50Ω/lead balanced.
Automatic Lead Compensation: <0.5% of span error.
RTD Sensor Current: 150mA (approximately).
Sensor Break
Protection:
Break detected within two seconds. Process Control outputs go to the
pre-set power value. High and Senor Break Alarms operate.
DC Linear Input
DC Linear Types & Ranges
Input
Type
Ranges Input
Type
Ranges
mA DC 0 to 20mA 4 to 20mA V DC 0 to 5V 1 to 5V
mV DC 0 to 50mV 10 to 50mV 2 to 10V 0 to 10V
DC Linear Performance
|